Magnetic Resonance Elastography and Supersonic Shear Imaging : simulation, experimental comparison and application to the characterization of the liver

Abstract : Elastography is an emerging medical imaging modality which permits to measure the mechanical properties of human soft tissue. The measured mechanical properties can serve as potential biomarkers for improving the management of diseases, from early diagnosis, to severity evaluation and therapy response monitoring. Among different approaches, Magnetic Resonance Elastography (MRE) and Supersonic Shear Imaging (SSI) have shown particular interests. The two modalities have been widely investigated for multiple clinical applications. However, each modality is challenged by specific acquisition and reconstruction conditions which may induce intra- and inter-modality measurement biases and hence impede the interchangeability of the two modalities. The first part of my thesis focused on identifying the measurement biases between MRE and SSI. Through a thorough methodological comparison study, we recognized different frequency characteristics of generated shear waves for the two modalities and modality specific reconstruction validity issues as the main sources for the measurement biases between the two modalities. Then through a dedicated simulation study, we established an in silico abaci to identify the favorable range of number of voxels per wavelength which leads to accurate and precise MRE. Moreover, resampling was proven effective to regulate poorly defined number of voxels per wavelength to the favorable range. The overall outcome, which is usually computed from the three acquired motionencoded directions, may further be improved by appropriate weighting strategies that are based on curl of shear displacement field. For SSI, we referred to the quality parameter provided by the manufacturer to reasonably eliminate unreliable results so as to further improve the measurement quality. After establishing the potential measurement biases between MRE and SSI, we incorporated the proposed quality optimization strategies into both modalities in order to perform unbiased experimental comparison studies between the two modalities. First, in vitro studies were carried out on commercial calibrated phantoms as well as home-made polyvinyl alcohol phantoms. Experimental results corroborate well the simulation findings. MRE and SSI measurements agree well witheach other when theory, experiment, and reconstruction biases are minimized. In vivo studies were then performed on the livers of two healthy volunteers. We found that when the liver is quasi-elastic, the quality-guided MRE and SSI measurements agree well with each other and hence are interchangeable. In case of viscoelastic liver tissue, both MRE and SSI measurements are frequency dependent. Thus frequency-specific measurements are essential for cross-validating the measurements of these two modalities.
Complete list of metadatas

Cited literature [136 references]  Display  Hide  Download

https://tel.archives-ouvertes.fr/tel-01682936
Contributor : Abes Star :  Contact
Submitted on : Friday, January 12, 2018 - 4:02:08 PM
Last modification on : Tuesday, September 29, 2020 - 4:46:50 AM
Long-term archiving on: : Sunday, May 6, 2018 - 2:50:03 AM

File

72398_YUE_2017_archivage.pdf
Version validated by the jury (STAR)

Identifiers

  • HAL Id : tel-01682936, version 1

Citation

Jinlong Yue. Magnetic Resonance Elastography and Supersonic Shear Imaging : simulation, experimental comparison and application to the characterization of the liver. Biological Physics [physics.bio-ph]. Université Paris Saclay (COmUE), 2017. English. ⟨NNT : 2017SACLS483⟩. ⟨tel-01682936⟩

Share

Metrics

Record views

760

Files downloads

316